ID | 116732 |
著者 |
Yoshihara, Hozumi
Tokushima University
|
キーワード | Digital
Intraoral radiography
PSP
CCD
|
資料タイプ |
学術雑誌論文
|
抄録 | The first digital sensor used in intraoral radiography, RadioVisioGraphy was introduced in 1987. It was based on the activity of a charged-Coupled Device (CCD) sensor. Later on, other digital sensors became commercially available, including the Digora featuring an imaging plate coated with photostimulable phosphor (PSP) in 1994 and the complementary metal-oxide semiconductor-based sensor, CDR active pixel sensor, in 1998. These technologies have recently undergone considerable improvement, and several dental clinics have switched from film-based systems to digital imaging. In the early days, digital sensors were considered inferior to film in clinical diagnostics, but according to recent research results, there is no longer any significant difference because of the improvement in performance. However, our latest research using a new evaluation method and a precise phantom model indicated some differences between the performance of CCD and PSP systems and further differences between PSP systems of different function and performance.
|
掲載誌名 |
Dental Health : Current Research
|
ISSN | 24700886
|
出版者 | SciTechnol
|
巻 | 4
|
号 | 2
|
開始ページ | 1000135
|
発行日 | 2018-05-15
|
権利情報 | This article is published under the terms of the Creative Commons Attribution (CC-BY) license (https://creativecommons.org/licenses/by/4.0/). It also allows immediate deposit of the final published version in other repositories without restriction on re-use.
|
EDB ID | |
出版社版DOI | |
出版社版URL | |
フルテキストファイル | |
言語 |
eng
|
著者版フラグ |
出版社版
|
部局 |
歯学系
|