至適透析や溶血回避のための血液ポンプ前の動脈圧の再評価 : 血液透析患者における血液ポンプ前の動脈圧モニタリングの重要性
Importance of pre pump arterial pres sure monitoring in hemodialysis patients
著者
柴田, 恵理子 徳島大学大学院医科学教育部(医学専攻) 徳島大学 教育研究者総覧 KAKEN研究者をさがす
長井, 幸二郎 The University of Tokushima KAKEN研究者をさがす
Takeuchi, Risa The University of Tokushima
Noda, Yasuhiro The University of Tokushima
Makino, Tomomi The University of Tokushima
Chikata, Yusuke The University of Tokushima
Hann, Michael Naval Medical Center San Diego
吉本, 咲耶 The University of Tokushima
小野, 広幸 The University of Tokushima
上田, 紗代 The University of Tokushima
村上, 太一 The University of Tokushima KAKEN研究者をさがす
松浦, 元一 The University of Tokushima KAKEN研究者をさがす
安部, 秀斉 The University of Tokushima KAKEN研究者をさがす
キーワード
Prepump arterial pressure
Hemolysis
pressure pillow
Dialysis
Blood flow
Vacuum monitor
資料タイプ
学位論文
抄録
Pre-pump arterial pressure (PreAP) is monitored to avoid generating excessive negative pressure. National Kidney Foundation K/DOQI clinical practice guidelines for vascular access recommend that PreAP should not fall below-250 mmHg because excessive negative PreAP can lead to a decrease in the delivery of blood flow, inadequate dialysis, and hemolysis. Nonetheless, these recommendations are consistently disregarded in clinical practice and pressure sensors are often removed from the dialysis circuit.Thus far,delivered blood flow has been reported to decrease at values more negative than -150 mmHg of PreAP. These values have been analyzed by an ultrasonic flowmeter and not directly measured. Furthermore, no known group has evaluated whether PreAP-induced hemolysis occurs at a particular threshold. Therefore, the aim of this study was to clarify the importance of PreAP in the prediction of inadequate dialysis and hemolysis. By using different diameter needles, human blood samples from healthy volunteers were circulated in a closed dialysis circuit. The relationship between PreAP and delivered blood flow or PreAP and hemolysis was investigated. We also investigated the optimal value for PreAP using several empirical monitoring methods, such as a pressure pillow. Our investigation indicated that PreAP is a critical factor in the determination of delivered blood flow and hemolysis,both of which occured at pressure values more negative than -150 mmHg. With the exception of direct pressure monitoring, commonly used monitoring methods for PreAP were determined to be ineffective. We propose that the use of a vacuum monitor would permit regular measurement of PreAP.
掲載誌名
Artificial Organs
ISSN
0160564X
15251594
cat書誌ID
AA00551950
AA11616446
出版者
International Center for Artificial Organs and Transplantation and Wiley Periodicals, Inc.
39
7
開始ページ
627
終了ページ
634
発行日
2015-05-01
備考
内容要旨・審査要旨・論文本文の公開
本論文は, 著者Eriko Shibataの学位論文として提出され, 学位審査・授与の対象となっている。
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng
著者版フラグ
博士論文全文を含む
文科省報告番号
甲第3117号
学位記番号
甲医第1352号
学位授与年月日
2017-10-26
学位名
博士(医学)
学位授与機関
徳島大学
部局
医学系
病院