ID | 105894 |
Title Alternative | 信頼性の高い持続的遊離皮弁モニタリングのための低温度下経皮的二酸化炭素分圧測定 : 動物実験と臨床経験
Flap Monitoring by TcPCO2
|
Author |
Abe, Yoshiro
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Hashimoto, Ichiro
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Goishi, Keiichi
The University of Tokushima
Kashiwagi, Keisuke
The University of Tokushima
Yamano, Masahiro
The University of Tokushima
Nakanishi, Hideki
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
|
Content Type |
Thesis or Dissertation
|
Description | Background: Measurement of transcutaneous oxygen pressure (TcPO2) and transcutaneous carbon dioxide pressure (TcPCO2) has been used for free flap monitoring. Because these values are obtained with sensor probes heated to 44°C, there is potential for low-temperature burns on skin flaps. We measured TcPO2 and TcPCO2 at 37°C in both animals and humans to determine the feasibility and safety of the procedure as a postoperative flap monitoring method.
Methods: Twelve epigastric island flaps were elevated in rabbits, and TcPO2 and TcPCO2 were measured at 37°C before and after ligation of the pedicles. In addition, TcPO2 and TcPCO2 at 37°C were measured in healthy men. Subsequently, the method was applied to postoperative monitoring of free flaps in 49 clinical cases. Results: TcPO2 and TcPCO2 values were significantly affected by the experimental flap elevation. A rapid increase in TcPCO2 was observed with both arterial and venous occlusion. In the healthy men, TcPO2 and TcPCO2 were measurable at all skin surface sites. In the clinical cases of free flap transfer, TcPO2 values remained very low for at least 72 hours. TcPCO2 values ranged from 40 to 70 mmHg for 72 hours in more than 80% of cases. In 2 cases, TcPCO2 values increased to more than 90 mmHg, and exploration surgery was performed. These compromised flaps were saved by reanastomosis of the veins. Conclusions: Continuous monitoring of TcPCO2 at 37°C can provide objective information and alert doctors and nurses to the need for checking the free flap. |
Journal Title |
Plastic and Reconstructive Surgery - Global Open
|
ISSN | 21697574
|
NCID | AA12786168
|
Publisher | Wolters Kluwer Health
|
Volume | 1
|
Issue | 2
|
Start Page | 1
|
End Page | 8
|
Published Date | 2013-05-13
|
Remark | 内容要旨・審査要旨・論文本文の公開:
内容要旨・審査要旨 : LID201310041001.pdf 内容要旨・審査要旨 : LID201406051002.pdf 本論文は, 著者Yoshiro Abeの学位論文として提出され, 学位審査・授与の対象となっている。 |
Rights | PRS GO uses the Creative Commons License Attribution-NonCommercial No Derivative 3.0 License (CCBY-NC-ND) for all articles and content published.( https://creativecommons.org/licenses/by-nc-nd/3.0/ )
|
EDB ID | |
DOI (Published Version) | |
URL ( Publisher's Version ) | |
FullText File | |
language |
eng
|
TextVersion |
ETD
|
MEXT report number | 甲第2622号
|
Diploma Number | 甲医第1167号
|
Granted Date | 2013-07-25
|
Degree Name |
Doctor of Medical Science
|
Grantor |
Tokushima University
|
departments |
Medical Sciences
|