ID | 110815 |
Title Alternative | Simple noise reduction for DWI
|
Author |
Konishi, Yuto
School of Health Sciences, Tokushima University
Kanazawa, Yuki
Institute of Biomedical Sciences, Tokushima University Graduate School
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Usuda, Takatoshi
School of Health Sciences, Tokushima University
Matsumoto, Yuki
School of Health Sciences, Tokushima University
Tokushima University Educator and Researcher Directory
Hayashi, Hiroaki
Institute of Biomedical Sciences, Tokushima University Graduate School
KAKEN Search Researchers
Matsuda, Tsuyoshi
MR Applications and Workflow Asia Pacific GE Healthcare Japan Corporation
Ueno, Junji
Institute of Biomedical Sciences, Tokushima University Graduate School
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Harada, Masafumi
Department of Radiology and Radiation Oncology, Institute of Biomedical Sciences, Tokushima University Graduate School
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
|
Keywords | Magnetic resonance imaging (MRI)
Diffusion weighted imaging
Probability distribution function (PDF)
Rician distribution
Gaussian distribution
Correction scheme
|
Content Type |
Journal Article
|
Description | Our purpose in this study was to reduce the noise in order to improve the SNR of Dw images with high b-value by using two correction schemes. This study was performed with use of phantoms made from water and sucrose at different concentrations, which were 10, 30, and 50 weight percent (wt%). In noise reduction for Dw imaging of the phantoms, we compared two correction schemes that are based on the Rician distribution and the Gaussian distribution. The highest error values for each concentration with use of the Rician distribution scheme were 7.3 % for 10 wt%, 2.4 % for 30 wt%, and 0.1 % for 50 wt%. The highest error values for each concentration with use of the Gaussian distribution scheme were 20.3 % for 10 wt%, 11.6 % for 30 wt%, and 3.4 % for 50 wt%. In Dw imaging, the noise reduction makes it possible to apply the correction scheme of Rician distribution.
|
Journal Title |
Radiological Physics and Technology
|
ISSN | 18650333
18650341
|
NCID | AA12236881
|
Publisher | Springer Japan
|
Volume | 9
|
Issue | 2
|
Start Page | 221
|
End Page | 226
|
Sort Key | 221
|
Published Date | 2016-03-16
|
Remark | Copyright © Japanese Society of Radiological Technology and Japan Society of Medical Physics 2016
The final publication is available at Springer via http://dx.doi.org/10.1007/s12194-016-0350-9. |
EDB ID | |
URL ( Publisher's Version ) | |
FullText File | |
language |
eng
|
TextVersion |
Author
|
departments |
Medical Sciences
|