ID | 111230 |
Author |
Hatakeyama, So
Kyoto University
Kunimura, Shinsuke
Kyoto University
Sasaki, Nobuharu
Kyoto University
Yamamoto, Takashi
Kyoto University
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Kawai, Jun
Kyoto University
|
Content Type |
Journal Article
|
Description | A portable instrument for two-dimensional X-ray fluorescence imaging was assembled with an X-ray source using a pyroelectric crystal, which was driven by a 9-V dry battery, a Si-PIN detector, a slit, and pulse motors. Line scanning for a mug and a knife-edge-scan of an iron sheet were carried out using this spectrometer. The sensitivity of the spectrometer was sufficient for elemental analysis of a mug using a 1 mm2 slit, and several elements, such as Co, Ni, Zn, Pb and Zr, were detected. The estimated spatial resolution using a 0.8-mm pinhole was 3.5 mm.
|
Journal Title |
Analytical Sciences
|
ISSN | 13482246
09106340
|
NCID | AA12022210
AA10500785
|
Publisher | The Japan Society for Analytical Chemistry
|
Volume | 24
|
Issue | 7
|
Start Page | 847
|
End Page | 850
|
Published Date | 2008-07-10
|
Rights | © 2008 by The Japan Society for Analytical Chemistry
|
EDB ID | |
DOI (Published Version) | |
URL ( Publisher's Version ) | |
FullText File | |
language |
eng
|
TextVersion |
Publisher
|
departments |
Science and Technology
|