Total for the last 12 months
number of access : ?
number of downloads : ?
ID 114307
Author
Nomura, Miharu Tokushima University
Shichijo, Maria Tokushima University
Keywords
Nitrite
seawater
flow injection analysis
standard addition method
Content Type
Journal Article
Description
A flow injection analysis of nitrite nitrogen (N-NO2-) in the samples which contain relatively high degrees of salt is described. The N-NO2- was optically detected by use of Griess method. Five pseudo-seawater samples containing 40 μg L-1 N-NO2- with 0, 1, 2, 3, or 4% sodium chloride (NaCl) were tested using two quantitative analysis approaches. In an absolute calibration method, approximately 30% of positive errors were found in the presence of NaCl. On the other hand, the errors fell within the range of −2.6 to +7.3% by introducing a standard addition method. The latter method has successfully been applied to the determination of nitrite nitrogen in real seawater and estuary water.
Journal Title
Journal of Flow Injection Analysis
ISSN
0911775X
24337374
NCID
AA10962565
Publisher
Japanese Association for Flow Injection Analysis
Volume
35
Issue
2
Start Page
59
End Page
61
Published Date
2018
EDB ID
DOI (Published Version)
URL ( Publisher's Version )
FullText File
language
eng
TextVersion
Publisher
departments
Pharmaceutical Sciences