Total for the last 12 months
number of access : ?
number of downloads : ?
ID 116864
Author
Watanabe, Hiroki Tokushima University
Sugiura, Jun Tokushima University
Kubáň, Petr CEITEC Masaryk University
Keywords
Ion-exchange membrane
electrolysis
ion chromatography
eluent generation
open tubular capillary column
Content Type
Journal Article
Description
We describe a miniaturized eluent generator (MEG), which can be integrated into open tubular capillary column ion chromatography (OTCC-IC). The MEG consists of one central eluent channel and two feed acid chambers, with each phase isolated by a cation-exchange membrane (CEM) and an anion-exchange membrane (AEM). Pure water flowed through the central eluent channel while acid solution (HA) was fed to both chambers. Applying the electric field to the MEG, H+ and A- simultaneously and respectively through the CEM and AEM to form HA in the central channel. The MEG successfully produced tartaric acid and 2,6-pyridinedicarboxylic acid as the eluent for cation OTCC-IC. The gradient elution improved the peak resolution and analytical window for five common cations (Na+, NH4+, K+, Ca2+, Mg2+).
Journal Title
Journal of Flow Injection Analysis
ISSN
0911775X
24337374
NCID
AA10962565
Publisher
Japanese Association for Flow Injection Analysis
Volume
38
Issue
1
Start Page
11
End Page
14
Published Date
2021
EDB ID
DOI (Published Version)
URL ( Publisher's Version )
FullText File
language
eng
TextVersion
Publisher
departments
Pharmaceutical Sciences