ID | 111244 |
Author |
Kawahito, Shinji
Tokushima University
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Takaishi, Kazumi
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Katayama, Toshiko
Tokushima University
Sakai, Yoko
Tokushima University
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Kawano, Hiroaki
Tokushima Prefectural Central Hospital
Matsuhisa, Munehide
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Shimada, Mitsuo
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Kitagawa, Tetsuya
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
Kitahata, Hiroshi
The University of Tokushima
Tokushima University Educator and Researcher Directory
KAKEN Search Researchers
|
Keywords | intensive insulin therapy
artificial endocrine pancreas
perioperative blood glucose management
continuous blood glucose monitoring
STG-55
|
Content Type |
Journal Article
|
Description | Background : Recent studies have shown that strict perioperative blood glucose management may reduce mortality and morbidity in critically ill adult patients. The purpose of this study was to assess the accuracy and efficacy of the intraoperative application of a newly developed, next-generation artificial endocrine pancreas (STG-55, Nikkiso Co., Ltd., Tokyo, Japan). Methods : Twenty patients scheduled to undergo surgery were enrolled in this study. The STG-55 is designed to be more user-friendly than its conventional counterpart (STG-22) while maintaining the latter’s fundamental functions, such as a closed-loop system using algorithms for insulin and glucose infusion. After anesthetic induction, a 20G intravenous catheter was inserted into a peripheral forearm vein and connected to a continuous blood glucose monitor. The resultant 105 scores for paired blood glucose values were compared by Bland-Altman analysis. Results : Stable blood glucose values were maintained automatically, and there were no complications related to use of the STG-55. A close correlation (r=0.96) was observed between continuous glucose measurements using the STG-55 and conventional intermittent glucose measurements. The difficulty of manipulation using this system was decreased by improved preparation procedures. Conclusion : The glycemic control system using the STG-55 could provide an alternative way to achieve effective and safe perioperative glycemic control.
|
Journal Title |
The Journal of Medical Investigation
|
ISSN | 13496867
13431420
|
NCID | AA11166929
AA12022913
|
Publisher | Faculty of Medicine Tokushima University
|
Volume | 62
|
Issue | 1-2
|
Start Page | 41
|
End Page | 44
|
Sort Key | 41
|
Published Date | 2015-02
|
EDB ID | |
DOI (Published Version) | |
URL ( Publisher's Version ) | |
FullText File | |
language |
eng
|
TextVersion |
Publisher
|
departments |
University Hospital
Medical Sciences
Institute of Advanced Medical Sciences
Oral Sciences
|