This study evaluated the effect of pitch on 256-slice helical computed tomography (CT) scans. Cylindrical water phantoms (CWP) were measured using axial and helical scans with various pitch values. The surface dose distributions of CWP were measured, and reconstructed images were obtained using filtered back-projection (FBP) and iterative model reconstruction (IMR). The image noise in each reconstructed image was decomposed into a baseline component and another component that varied along the z-axis. The baseline component of the image noise was highest at the center of the reconstructed image and decreased toward the edges. The normalized 2D power spectra for each pitch were almost identically distributed. Furthermore, the ratios of the 2D power spectra for IMR and FBP at different pitch values were obtained. The magnitudes of the components varying along the z-axis were smallest at the center of the reconstructed image and increased toward the edge. The ratios of the 3D power spectra on the fx axis for IMR and FBP at different pitch values were obtained. The results showed that the effect of the pitch was related to the component that varied along the z-axis. Furthermore, the pitch had a smaller effect on IMR than on FBP.
Journal Title
Sensors
ISSN
14248220
Publisher
MDPI
Volume
23
Issue
7
Start Page
3472
Published Date
2023-03-26
Rights
© 2023 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access article distributed under the terms and conditions of the Creative Commons Attribution (CC BY) license (https://creativecommons.org/licenses/by/4.0/).
EDB ID
DOI (Published Version)
URL ( Publisher's Version )
FullText File
language
eng
TextVersion
Publisher
departments
Science and Technology
Medical Sciences