ID | 109851 |
タイトルヨミ | シカ キョウセイヨウ アンカー スクリュー マイニュウジ ノ アンテイセイ ニ カンスル セイタイリキガクテキ ケンキュウ
|
タイトル別表記 | Stress Analysis on Surrounding Bone of Anchor Screw during Orthodontic Force Application
|
著者 |
井上, 雅秀
徳島大学大学院口腔科学教育部口腔顎顔面矯正学分野
黒田, 晋吾
徳島大学大学院ヘルスバイオサイエンス研究部口腔顎顔面矯正学分野
|
キーワード | アンカースクリュー
有限要素解析
傾斜埋入
|
資料タイプ |
学術雑誌論文
|
抄録 | Objective: The purpose of this study was to evaluate the influence of placement angle and force direction on the primary stability of orthodontic miniscrews by analyzing a novel threedimensional finite element model (FEM) approximating the interface between the screw and surrounding bone.
Materials and Methods: Three-dimensional finite element models were made with 6-mm-long miniscrews in diameters of 1.2 mm. Four insertion angles ranging from 0̊ (perpendicular to the bone surface) to 45̊ were examined. A load with 2 N was applied to the center of the screw head in four directions (upward, downward and on the right and left sides). Results: For miniscrews at the same insertion angle, the stress was the highest (or lowest) under the downward (or upward) force condition. The stress increased as the insertion angle increased, except under the upward force condition. An analysis of the stress distribution in the surrounding bone showed that the most of the stress was absorbed in the cortical bone. Conclusions: The stress distribution around the miniscrew and surrounding bone during the application of orthodontic force is closely related to the force direction, and insertion angle. |
掲載誌名 |
四国歯学会雑誌
|
ISSN | 09146091
|
cat書誌ID | AN10050046
|
出版者 | 四国歯学会
|
巻 | 27
|
号 | 1
|
開始ページ | 27
|
終了ページ | 34
|
並び順 | 27
|
発行日 | 2014-06-30
|
EDB ID | |
フルテキストファイル | |
言語 |
jpn
|
著者版フラグ |
出版社版
|
部局 |
歯学系
|