artificial pancreas
continuous glucose monitoring
euglycemic hyperinsulinemic clamp study
hyperglycemic clamp study
Parkes consensus error grid
資料タイプ
学術雑誌論文
抄録
Background:
Glucose values of continuous glucose monitoring (CGM) have time delays compared with plasma glucose (PG) values. Artificial pancreas (STG-55, Nikkiso, Japan) (AP), which measures venous blood glucose directly, also has a time delay because of the long tubing lines from sampling vessel to the glucose sensor. We investigate accuracy and time delay of CGM and AP in comparison with PG values during 2-step glucose clamp study.
Methods:
Seven patients with type 2 diabetes and 2 healthy volunteers were included in this study. CGM (Enlite sensor, Medtronic, CA) was attached on the day before the experiment. Hyperglycemic (200 mg/dL) clamp was performed for 90 minutes, followed by euglycemic (100 mg/dL) hyperinsulinemic (100 μU/mL) clamp for 90-120 minutes using AP. CGM sensor glucose was calibrated just before and after the clamp study. AP and CGM values were compared with PG values.
Results:
AP values were significantly lower than PG values at 5, 30 minute during hyperglycemic clamp. In comparison, CGM value at 0 minute was significantly higher, and its following values were almost significantly lower than PG values. The time delay of AP and CGM values to reach maximum glucose levels were 5.0 ± 22.3 (NS) and 28.6 ± 32.5 (p<0.05) min, respectively. Mean absolute rate difference of CGM was significantly higher than AP (24.0 ± 7.6 vs. 15.3 ± 4.6, p < 0.05) during glucose rising period (0-45 min), however, there are no significant difference during other periods.
Conclusions:
Both CGM and AP failed to follow plasma glucose values during non-physiologically rapid glucose rising, however, indicated accurate values during physiological glucose change.
掲載誌名
Journal of Diabetes Science and Technology
ISSN
19322968
出版者
Diabetes Technology Society|SAGE Publications
11
6
開始ページ
1096
終了ページ
1100
発行日
2017-10-09
権利情報
Akio Kuroda, Satoshi Taniguchi, Yuko Akehi, Hiroyasu Mori, Motoyuki Tamaki, Reiko Suzuki, Yinhua Otsuka, Munehide Matsuhisa, Accuracy and Time Delay of Glucose Measurements of Continuous Glucose Monitoring and Bedside Artificial Pancreas During Hyperglycemic and Euglycemic Hyperinsulinemic Glucose Clamp Study, Journal of Diabetes Science and Technology (Volume: 11 issue: 6) pp.1096-1100. Copyright © 2017 Diabetes Technology Society. DOI: 10.1177/1932296817735122.
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng
著者版フラグ
著者版
部局
先端酵素学研究所