直近一年間の累計
アクセス数 : ?
ダウンロード数 : ?
ID 115426
タイトル別表記
ICG fluorescence catheter system in TaTME
著者
太田, 昇吾 The University of Tokushima
東島, 潤 The University of Tokushima KAKEN研究者をさがす
宮谷, 知彦 The University of Tokushima KAKEN研究者をさがす
石川, 大地 The University of Tokushima
キーワード
TaTME
ICG
fluorescence
urethral injury
rectal cancer
資料タイプ
学術雑誌論文
抄録
Background. Sometimes intraoperative urethral injury occurs in Transanal total mesorectal excision (TaTME). The aim of this study is to investigate the usefulness of indocyanine green (ICG) fluorescent catheter system for avoiding intraoperative urethral injury in TaTME in experimental model. Methods. A urethral catheter was filled with the mixture of albumin and ICG and raw hams were applied in layers as the surrogate model of rectourethral muscle. The detectability of ICG fluorescence in this catheter was investigated by using laparoscope-type fluorescence camera system. Results. Fluorescence was detected when ICG was mixed with albumin or peripheral blood. ICG fluorescence could be detected within 4 mm depth of layered raw hams as the surrogate model. Quantitative analysis of the picture detected that ICG fluorescence plateaued in lower concentration than that of serum. Conclusion. ICG fluorescent catheter system may be useful for avoiding intraoperative urethral injury in TaTME.
掲載誌名
The Journal of Medical Investigation
ISSN
13496867
13431420
cat書誌ID
AA11166929
出版者
Tokushima University Faculty of Medicine
67
3-4
開始ページ
285
終了ページ
288
並び順
285
発行日
2020-08
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng
著者版フラグ
出版社版
部局
医学系
病院