キーワード | Amplitude modulated multiplexed flow analysis
amplitude modulation
internal standard method
Fourier transform
1,10-phenanthroline
Fe(II) ion
|
資料タイプ |
学術雑誌論文
|
抄録 | A new concept of flow analysis, internal standard-amplitude modulated multiplexed flow analysis, is proposed. A proof of concept was verified by applying it to the determination of ferrous ion (Fe2+) by 1,10-phenanthroline (o-Phen) spectrophotometry. The flow rates of sample solutions containing Methylene Blue (MB) as an internal standard substance were sinusoidally varied at different frequencies. The solutions were merged with a color reagent (o-Phen) solution, while the total flow rate was held constant. Downstream, analytical signals were monitored at the maximum absorption wavelengths of Fe2+-o-Phen complex and of MB (510 and 644 nm, respectively). The signals were respectively analyzed by fast Fourier transform. The concentrations of the analytes in respective samples were simultaneously determined from the amplitudes of the corresponding wave components. The precision, linearity of the calibration curve, limit of detection and robustness against deliberate fluctuation in flow rate were greatly improved by introducing the internal standard method. Good recoveries of around 100% were obtained for Fe2+ spiked into real water samples.
|
掲載誌名 |
Analytical Sciences
|
ISSN | 13482246
09106340
|
cat書誌ID | AA10500785
|
出版者 | The Japan Society for Analytical Chemistry
|
巻 | 33
|
号 | 12
|
開始ページ | 1363
|
終了ページ | 1368
|
発行日 | 2017-12-10
|
EDB ID | |
出版社版DOI | |
出版社版URL | |
フルテキストファイル | |
言語 |
eng
|
著者版フラグ |
出版社版
|
部局 |
薬学系
|