ID | 70726 |
著者 |
岡﨑, 憲二
Department of Thoracic, Endocrine Surgery and Oncology, Institute of Health Biosciences, the University of Tokushima Graduate School
丹黒, 章
Department of Thoracic, Endocrine Surgery and Oncology, Institute of Health Biosciences, the University of Tokushima Graduate School
徳島大学 教育研究者総覧
KAKEN研究者をさがす
森本, 忠興
Department of Thoracic, Endocrine Surgery and Oncology, Institute of Health Biosciences, the University of Tokushima Graduate School|Department of Surgery, Shikoku Central Hospital of the Mutual Aid
徳島大学 教育研究者総覧
KAKEN研究者をさがす
小谷, 亮介
Graduate School of Advanced Technology and Science, the University of Tokushima
服部, 圭悟
Graduate School of Advanced Technology and Science, the University of Tokushima
安野, 恵実子
Department of Systems and Control Engineering,Anan National College of Technology
芥川, 正武
Department of Electrical and Electronic Engineering, Institute of Technology and Science, the University of Tokushima Graduate School
徳島大学 教育研究者総覧
KAKEN研究者をさがす
木内, 陽介
Department of Electrical and Electronic Engineering, Institute of Technology and Science, the University of Tokushima Graduate School
徳島大学 教育研究者総覧
KAKEN研究者をさがす
|
キーワード | noninvasive
local tissue
electrical impedance tomography
measurement
breast diseases
|
資料タイプ |
学術雑誌論文
|
抄録 | The objective of this study is to develop a device for noninvasive local tissue electrical impedance tomography (EIT) using divided electrodes with guard electrodes and to validate its effectiveness using bioequivalent phantoms. For this purpose, we prepared a measurement device and bioequivalent phantoms, measured the electrical characteristics of the phantoms, and validated the method using the phantoms. Monolayer phantoms mimicking the brain and muscle and bilayer phantoms consisting of muscle and brain layers were prepared. The relative differences between the measured electrical conductivities of the monolayer brain and muscle phantoms and the true values determined by the 4-electrode method were both less than 10%. The relative differences between the measured and true values in the bilayer phantoms were less than 20% in both layers. The biological impedance measurement device that we developed was confirmed to be effective for impedance measurement in bilayer phantoms with different electrical impedances. To develop a device for the early diagnosis of breast diseases, the development of a multi-layer phantom and demonstration of the effectiveness of the device for its examination are necessary. If the device that we developed makes impedance measurement in breast tumors possible, it may be used as a new diagnostic modality for breast diseases.
|
掲載誌名 |
The journal of medical investigation : JMI
|
ISSN | 13431420
|
cat書誌ID | AA11166929
|
巻 | 57
|
号 | 3-4
|
開始ページ | 205
|
終了ページ | 218
|
並び順 | 205
|
発行日 | 2010-08
|
備考 | The journal of medical investigation : http://medical.med.tokushima-u.ac.jp/jmi/index.html
|