ID 114439
著者
Cui, Minchao Xi’an Jiaotong University|Tokushima University
出口, 祥啓 Tokushima University|Xi’an Jiaotong University 徳島大学 教育研究者総覧 KAKEN研究者をさがす
Wang, Zhenzhen Tokushima University|Xi’an Jiaotong University
Tanaka, Seiya Tokushima University
Fujita, Yuki Tokushima University
Zhao, Shengdun Xi’an Jiaotong University
キーワード
long-short DP-LIBS
steel
manganese
improvement
analytical performance
quantitative analysis
資料タイプ
学術雑誌論文
抄録
A long-short double pulse laser-induced breakdown spectroscopy (long-short DP-LIBS) method was employed to improve the performance of LIBS for the measurement of manganese in steel samples. The long pulse was generated by a Nd:YAG laser which was operated at free runing (FR) mode. To investigate the detection ability without sample preparation, the steel washers were tested using SP-LIBS and long-short DP-LIBS, respectively. The measurement results show that the long-short DP-LIBS method was able to record clear spectra from the rusty steel washers. The steel washers were also measured after the polishing process. The measurement results show that the signal intensity was enhanced by long-short DP-LIBS. Through the observation with scanning electron microscope (SEM) on the laser craters, the results suggest that the improvement of detection ability can be attributed to the pre-radiation effect of long-pulse laser beam. Next, the analytical performance for quantitative measurement of manganese was evaluated by employing ten standard steel samples. The results show that the linearty fit (R2) of calibration curve is 0.988 for long-short DP-LIBS, whereas, R2 is only 0.810 for SP-LIBS under the same measurement conditions. The five times repeated measurement results show that the average Relative Standard Deviation (RSD) of the tested samples is 29.3% for SP-LIBS and is 10.5% for long-short DP-LIBS. The prediction results also show that the average Relative Error of Prediction (REP) is 94.9% for SP-LIBS and it 4.9% for long-short DP-LIBS. The experimental results in current work demonstrate that long-short DP-LIBS is promising for the on line measurement of steel in the steelmaking plant.
掲載誌名
Applied Spectroscopy
ISSN
00037028
19433530
cat書誌ID
AA00543544
出版者
Society for Applied Spectroscopy|SAGE Publishing
73
2
開始ページ
152
終了ページ
162
発行日
2018-09-28
権利情報
Minchao Cui, Yoshihiro Deguchi, Zhenzhen Wang, Seiya Tanaka, Yuki Fujita, Shengdun Zhao, Improved Analysis of Manganese in Steel Samples Using Collinear Long–Short Double Pulse Laser-Induced Breakdown Spectroscopy (LIBS), Applied Spectroscopy (Volume: 73 issue: 2) pp. 152-162 . Copyright ©2018 SAGE Publications. DOI: 10.1177/0003702818803943.
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng