直近一年間の累計
アクセス数 : ?
ダウンロード数 : ?
ID 119009
著者
Kakiuchi, Naoya Tokushima University
キーワード
Feedback control
Feedback-based flow ratiometry
Flow titration
Process control
Online analysis
SDGs
資料タイプ
学術雑誌論文
抄録
The concept of an integrated automated continuous flow method with dual feedback controls is presented for diluting a stock solution to provide a solution of a given concentration. The one control is used for the online process monitoring by a feedback-based flow ratiometry, where the product (the diluted liquid) is titrated through the rapid bidirectional scan of the product/reagent flow ratio. The feedback control limits the scanning to the necessary range to increase the analytical throughput. The other control is used for the process control to output the product with a preset concentration. The merging ratio of the stock solution and a solvent (diluent) is changed based on the information from the online analysis. The concept was verified by applying it to producing 0.1 mol dm−3 CH3COOH. When the stock concentration was changed from 0.1 (reference concentration) to 0.3 and then 0.2 mol dm−3, the system searched for the suitable merging ratio and converged the output concentration to the reference value within 7.43 min with a relative error below 1.05 %. The mean throughput rate of the process analysis was 11.2 titrations min−1. Successful results were also obtained for the 0.1 mol dm−3 HCl production. The present concept could be the basis for process control with reduced wasteful output and effluent treatment with eco-friendly treated water discharge, resulting in the contribution to SDGs’ goals of 6 (Clean water and sanitation), 9 (Industry, innovation and infrastructure), and 14 (Life below water).
掲載誌名
Analytical Sciences
ISSN
13482246
09106340
cat書誌ID
AA10500785
出版者
The Japan Society for Analytical Chemistry|Springer Nature
39
5
開始ページ
755
終了ページ
759
発行日
2022-10-23
備考
This version of the article has been accepted for publication, after peer review (when applicable) and is subject to Springer Nature’s AM terms of use (https://www.springernature.com/gp/open-research/policies/accepted-manuscript-terms), but is not the Version of Record and does not reflect post-acceptance improvements, or any corrections. The Version of Record is available online at: https://doi.org/10.1007/s44211-022-00206-3
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng
著者版フラグ
著者版
部局
薬学系