Failure of Hem-o-lok clips
著者
Yanagi, Masato Nippon Medical School
Hamasaki, Tsutomu Nippon Medical School
Shimizu, Hiroyuki Hasegawa Hospital
Kimata, Ryoji Nippon Medical School
Takeda, Hayato Nippon Medical School
Akatsuka, Jun Nippon Medical School
Endo, Yuki Nippon Medical School
Hayashi, Tatsuro Nippon Medical School
Kondo, Yukihiro Nippon Medical School
キーワード
Hem-o-lok clip
laparoscopic renal surgery
renal artery
資料タイプ
学術雑誌論文
抄録
Hem-o-lok clips are commonly used for renal artery ligation in laparoscopic renal surgery. However, failure of the renal artery ligation clips is potentially fatal. A 61-year-old man underwent hand-assisted laparoscopic nephroureterectomy using a retroperitoneal approach for left ureteral carcinoma. One hour postoperatively, he was diagnosed with hemorrhagic shock. An immediate laparotomy revealed two closed, undamaged Hem-o-lok clips around the left renal artery. Pulsatile bleeding was observed, and the renal artery was immediately ligated with non-absorbable thread. We determined that the failure of the Hem-o-lok clips on the renal artery was caused by the lack of space between the two Hem-o-lok clips and the distal renal artery cuff beyond the distal clip. To prevent a potentially fatal failure of the renal artery ligation clips, one should maintain a sufficient space between the Hem-o-lok clips and an adequate distal renal artery cuff beyond the distal clip.
掲載誌名
The Journal of Medical Investigation
ISSN
13496867
13431420
cat書誌ID
AA11166929
出版者
Tokushima University Faculty of Medicine
68
3-4
開始ページ
393
終了ページ