ID 115925
著者
Asahara, Takashi Kanazawa University|Okayama University
Goto, Sota Kanazawa University
Kimoto, Natsumi Kanazawa University
Takegami, Kazuki Kanazawa University
Maeda, Tatsuya Kanazawa University
Okazaki, Tohru Nagase Landauer
Hashizume, Takuya Nagase Landauer
キーワード
calibration factor
eye lens dose
IVR
OSL dosimeter
scattered x-ray
資料タイプ
学術雑誌論文
抄録
The eye lens is a sensitive organ of which an x‐ray exposure dose should be managed during interventional radiology (IVR). In the actual situations, the eye lens is exposed to scattered x‐rays; they have different from the standard x‐ray energies which are used for general dose calibration of the dosimeter. To perform precise dose measurement, the energy dependence of the dosimeter should be properly accounted for when calibrating the dosimeter. The vendor supplies a calibration factor using 80‐kV diagnostic x‐rays under a free‐air condition. However, whether it is possible to use this calibration factor to evaluate the air kerma during the evaluation of the eye lens dose is unclear. In this paper, we aim to precisely determine calibration factors, and also examine the possible application of using a vendor‐supplied calibration factor. First, the x‐ray spectrum at the eye lens position during fluoroscopy was measured with a CdTe x‐ray spectrometer. We mimicked transfemoral cardiac catheterization using a human‐type phantom. Second, we evaluated the doses and calibration factors at three dosimetric points: front and back of protective goggles, and the front of the head (eye lens position). We used the measured x‐ray spectrum to determine the incident photon distribution in the eye lens regions, and x‐ray spectra corresponding to the dosimetric points around the eye lens were estimated using Monte Carlo simulation. Although the calibration factors varied with dosimetric positions, we found that the factors obtained were similar to the vendor‐supplied calibration factor. Furthermore, based on the experiment, we propose a practical way to calibrate an OSL dosimeter in an actual clinical situation. A person evaluating doses can use a vendor‐supplied calibration factor without any corrections for energy dependences, only when they add a systematic uncertainty of 5%. This evidence will strongly support actual exposure dose measurement during a clinical study.
掲載誌名
Journal of Applied Clinical Medical Physics
ISSN
15269914
cat書誌ID
AA12034060
出版者
Wiley Periodicals|American Association of Physicists in Medicine
21
11
開始ページ
263
終了ページ
271
発行日
2020-10-14
権利情報
This is an open access article under the terms of the Creative Commons Attribution License(https://creativecommons.org/licenses/by/4.0/), which permits use, distribution and reproduction in any medium, provided the original work is properly cited.
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng
著者版フラグ
出版社版
部局
医学系