ID | 115180 |
タイトル別表記 | Axial Dispersion-Reduced Cation Exchange Module and Its Application to Ion Chromatography
|
著者 |
岡本, 和将
徳島大学
岡部, 芹香
徳島大学
|
キーワード | ion chromatography
anion
suppressor
Nafion tube
axial dispersion
|
資料タイプ |
学術雑誌論文
|
抄録 | The suppressor used in ion chromatography is a module placed between the separation column and conductivity detector. The axial dispersion of the sample solution in the suppressor causes a decline of detection sensitivity due to the broadening of the detection peaks. In this paper, we describe a monofilament inserted Nafion tube-type suppressor for anion analysis, which has an axial dispersion-reduced structure. A standard solution containing fluoride, formate, chloride, nitrite, nitrate, and sulfate was measured by means of an ion chromatograph with the fabricated suppressor. The number of theoretical plates and resolution were, respectively, 1.6 to 2.9 times and 1.3 to 1.8 times higher for the fabricated suppressor than a commercially available electrolytic suppressor. These results suggest that the fabricated suppressor has less axial dispersion of the sample than the electrolytic suppressor. In addition, the background level of the conductivity signal with the fabricated suppressor was suppressed to almost the same level as the electrolytic suppressor.
|
掲載誌名 |
Journal of Flow Injection Analysis
|
ISSN | 0911775X
24337374
|
cat書誌ID | AA10962565
|
出版者 | Japanese Association for Flow Injection Analysis
|
巻 | 37
|
号 | 1
|
開始ページ | 3
|
終了ページ | 8
|
発行日 | 2020
|
EDB ID | |
出版社版DOI | |
出版社版URL | |
フルテキストファイル | |
言語 |
jpn
|
著者版フラグ |
出版社版
|
部局 |
薬学系
|