ID | 115361 |
タイトル別表記 | Assessment of accidental events in images during capsule endoscopy
カプセル内視鏡における画像不良
|
著者 |
藤井, 有美子
徳島大学
小林, 誠司
徳島大学
大西, 芳明
徳島大学
|
キーワード | Capsule endoscopy
Image defect
Electromagnetic interference
|
資料タイプ |
学術雑誌論文
|
抄録 | Recently, there have been some reports of electromagnetic interference during capsule endoscopy(CE)in regards to other medical devices and facility environment. These accidental events have not been analyzed sufficiently although academic societies and related ministries are proposing guidelines with cautious attention. Based on this, detailed analyses and development of solution are required. In this study, we review our experience with cases of accidental events during CE and assessed the causes and preventive solutions. A total of149CE cases(male :89, female :60, mean age :56±18.8)from January 2017 until December 2019 were analyzed retrospectively. Four cases(2%)with defective images were noted. Detailed events were as follows ;1 disorder of the sensor array, 2 recorder freezes, and 1 electromagnetic interference. These mechanical and electromagnetic troubles should be assessed and shared among medical staffs and manufacturers to propose possible solutions and perform CE safely.
|
掲載誌名 |
四国医学雑誌
|
ISSN | 00373699
|
cat書誌ID | AN00102041
|
出版者 | 徳島医学会
|
巻 | 76
|
号 | 3-4
|
開始ページ | 159
|
終了ページ | 164
|
並び順 | 159
|
発行日 | 2020-08-25
|
フルテキストファイル | |
言語 |
jpn
|
著者版フラグ |
出版社版
|
部局 |
医学系
|