ID | 119475 |
タイトル別表記 | Mismatch between AR navigation and tumor
|
著者 |
藤谷, 順三
Tokushima University|Sapporo Medical University
Ogawa, Takahiro
Tokushima University|Gifu University
|
キーワード | augmented reality
spinal tumor surgery
microscope-based navigation
pitfall
|
資料タイプ |
学術雑誌論文
|
抄録 | Background : Augmented reality navigation is the one of the navigation technologies that allows computer-generated virtual images to be projected onto a real-world environment. Augmented reality navigation can be used in spinal tumor surgery. However, it is unknown if there are any pitfalls when using this technique. Case presentation : The patient in this report underwent complete resection of a cauda equina tumor at the L2–L3 level using microscope-based augmented reality navigation. Although the registration error of navigation was < 1 mm, we found a discrepancy between the augmented reality navigation images and the actual location of the tumor, which we have called “navigation mismatch”. This mismatch, which was caused by the mobility of the spinal tumor in the dura mater, seems to be one of the pitfalls of augmented reality navigation for spinal tumors. Conclusions : Combined use of intraoperative ultrasound and augmented reality navigation seems advisable in such cases.
|
掲載誌名 |
The Journal of Medical Investigation
|
ISSN | 13496867
13431420
|
cat書誌ID | AA11166929
|
出版者 | Tokushima University Faculty of Medicine
|
巻 | 71
|
号 | 1-2
|
開始ページ | 174
|
終了ページ | 176
|
並び順 | 174
|
発行日 | 2024-02
|
EDB ID | |
出版社版DOI | |
出版社版URL | |
フルテキストファイル | |
言語 |
eng
|
著者版フラグ |
出版社版
|
部局 |
医学系
病院
|