ID | 111235 |
タイトル別表記 | Analysis of Valence for Chromium in Soil and Plastic Samples Using Laboratory XAFS Spectrometer
|
著者 |
宮内, 宏哉
中小企業技術センター|京都大学
中西, 貞博
中小企業技術センター
河合, 潤
京都大学
|
キーワード | laboratory XAFS
XANES
hexavalent chromium
soil
plastic
|
資料タイプ |
学術雑誌論文
|
抄録 | Hexavalent {Cr(VI)} and trivalent chromium {Cr(III)} in environmental materials such as soil and plastics were investigated using a laboratory X-ray absorption fine structure (XAFS) spectrometer equipped a W anode X-ray tube. The Cr-K X-ray absorption edges were observed on the XANES spectra of soil (total chromium 606 ppm) and plastic (213 ppm) samples with a fluorescence mode, while the signal-to-noise (SN) ratios of these spectra were too low to determine the height of pre-edge peak of Cr(VI) and the value of the energy on the Cr-K X-ray absorption edge. The XANES spectra of reference samples prepared by Cr2O3 {Cr(III)} and CrO3 {Cr(VI)} particle reagents were also recorded with the transmission mode. In order to improve the SN ratios of XANES spectra, a smoothing treatment by Savitzky-Golay method was applied to the XANES spectra of the soil and plastic samples. By the smoothing treatment, SN ratios of these spectra were improved, though the height of pre-edge peak on the XANES spectra of Cr(VI) reference sample had decreased less than 5%. The height of pre-edge peak and the value of the energy on the Cr-K X-ray absorption edge of soil and plastic samples were given by smoothed XANES spectra. Using the height of pre-edge peak and the value of the energy on the Cr-K X-ray absorption edge of reference samples, the ratios of Cr(VI)/Cr(III) on the soil and plastic samples were calculated.
|
掲載誌名 |
分析化学
|
ISSN | 05251931
|
cat書誌ID | AN00222633
AA11681958
|
出版者 | 日本分析化学会
|
巻 | 58
|
号 | 4
|
開始ページ | 321
|
終了ページ | 325
|
発行日 | 2009-05-04
|
権利情報 | © The Japan Society for Analytical Chemistry 2009
|
EDB ID | |
出版社版DOI | |
出版社版URL | |
フルテキストファイル | |
言語 |
jpn
|
著者版フラグ |
出版社版
|
部局 |
理工学系
|