ID | 111620 |
著者 |
Hayashi, Shigehito
The University of Tokushima
Shibata, Hiroshi
The University of Tokushima
|
キーワード | contrast medium
iomeprol
hemofiltration
hemodiafiltration
|
資料タイプ |
学術雑誌論文
|
抄録 | The contrast medium-removing effect of hemofiltration (HF) and hemodiafiltration (HDF) was experimentally investigated using a bovine blood tank model. HF and HDF were performed at a blood flow rate of 100ml/min with a polysulfone hemofilter (PS filter-CF ; membrane area : 0.7 m2). Two hundred milliliters of iomeprol (300 mgI/ml) was administered by a single injection into 4 liters of bovine blood. The blood half-lives of iomeprol were 1.0 hr for the high flow rate HDF group [replacement fluid flow rate (QF) : 10 ml/min and dialysate flow rate (QD) : 40 ml/min], 1.8 hr for the HDF group (QF : 10 ml/min and QD : 10 ml/min), and 3.8 hr for the HF group (QF : 10 ml/min). The mean clearance rates were 39.7 ml/min for the high flow rate HDF group, 21.4 ml/min for the HDF group, and 12.0 ml/min for the HF group. Iomeprol was mostly excreted in the waste fluid. It is concluded that HDF can remove contrast media more effectively than HF.
|
掲載誌名 |
The Journal of Medical Investigation
|
ISSN | 13431420
|
cat書誌ID | AA11166929
|
出版者 | The University of Tokushima School of Medicine
|
巻 | 45
|
号 | 1-4
|
開始ページ | 87
|
終了ページ | 93
|
並び順 | 87
|
発行日 | 1998-08
|
EDB ID | |
出版社版URL | |
フルテキストファイル | |
言語 |
eng
|
著者版フラグ |
出版社版
|
部局 |
医学系
|