直近一年間の累計
アクセス数 : ?
ダウンロード数 : ?
ID 113648
著者
Isayama, Takaaki Tokushima University
キーワード
Dose area product (DAP) meter
Observer test
Sign test
Wilcoxon test
Significant difference
Small electrode
資料タイプ
学術雑誌論文
抄録
The purpose of this study is to propose a new method of image evaluation using statistical analysis. We used the Sign test and the Wilcoxon test to analyze the statistical significance of image differences. Using this method, we evaluated whether the small electrode of the DAP meter appears in the X-ray image. Two observed values, which were obtained by averaging all values under all exposure conditions, were compared. All the observation tests showed the same sign. Thus, the results proved that the small electrode of the DAP meter is not present on the image. Using this method, it became possible to prove that the electrode was not depicted, which was impossible to determine using conventional methods. The method combining both the Sign test and the Wilcoxon test can be useful in image evaluation.
掲載誌名
Radiological Physics and Technology
ISSN
18650333
18650341
cat書誌ID
AA12236881
出版者
Japanese Society of Radiological Technology|Japan Society of Medical Physics|Springer Singapore
12
2
開始ページ
156
終了ページ
160
発行日
2019-03-11
備考
The final publication is available at link.springer.com.
EDB ID
出版社版DOI
出版社版URL
フルテキストファイル
言語
eng
著者版フラグ
著者版
部局
医学系