ID | 117203 |
タイトル別表記 | Monitoring of Bacterial-contamination of Dental Unit Water Lines using ATP-bioluminescence
Bacterial-contamination Monitoring
|
著者 |
Yokota, Kenji
Okayama University
Kokeguchi, Susumu
Okayama University
|
キーワード | ATP bioluminescence
Dental unit water line
Bacteria
Contamination
Infection control
ATP-bioluminescence analysis
Bacteria contamination
|
資料タイプ |
学術雑誌論文
|
抄録 | Bacterial contamination of dental unit waterlines (DUWLs) was evaluated by using ATP-bioluminescence analysis and conventional culture method. Water samples (n=44) from DUWLs were investigated for heterotrophic bacteria by culture on R2A agar, which ranged from 1.4×103 to 2.7×105 CFU/mL. The ATP-bioluminescence results for DUWL samples were ranged from 6 to 1189 RLU and obtained within one minutes. These results were well correlated with the culture results (r=0.727-0.855). We conclude that differences in the bacterial contamination of each water supply were confirmed by the ATP-bioluminescence assay. This method would be potentially useful for rapid and simple monitoring of DUWL bacterial contamination.
|
掲載誌名 |
Journal of Hospital Infection
|
ISSN | 01956701
|
cat書誌ID | AA00258209
AA11538375
|
出版者 | The Healthcare Infection Society|Elsevier
|
巻 | 94
|
号 | 4
|
開始ページ | 393
|
終了ページ | 396
|
発行日 | 2016-08-06
|
権利情報 | © 2016. This manuscript version is made available under the CC-BY-NC-ND 4.0 license http://creativecommons.org/licenses/by-nc-nd/4.0/
|
EDB ID | |
出版社版DOI | |
出版社版URL | |
フルテキストファイル | |
言語 |
eng
|
著者版フラグ |
著者版
|
部局 |
歯学系
|